Introduction to Technical Parameters of Standard Group Tracking Tester

Author:QINSUN Released in:2023-10 Click:90

Technical parameters of tracking tester:
1. Electrode material: electrode tip - platinum metal (platinum)
2. Electrode pressure: 1.0 ± 0.05 N
3. Electrode distance: 4.0 mm ± 0.01 mm, included angle 60°±5
4. Electrode voltage: 100 ~ 600 V (48 ~ 60 Hz) adjustable, the voltage drop is less than 10% when the short circuit current is 1.0 ± 0.1 A
5. Drop height: 30 ~ 40mm adjustable
6. Drop size: 44~55 drops/1 cm3 adjustable
7. Falling time: 30 s ± 5 s
8. Number of liquid drops: 0 ~ 9999 times, adjustable
9. Leakage judgment: The loop current is greater than 0. If the loop current is greater than 0 and maintained for 2 seconds, the relay will act to cut off the current, indicating that the sample is not qualified
10 . Testing Process: Automatic control of the test program. Independent ventilation
11. Reference standards: IEC60112, GB/T4207, GB4706.1
12. Studio volume: 0.1 m3 (can be customized to 0.5 m3 or 0.75 m3)
13. External dimension of the equipment: 630 mm wide × 390 mm deep IEC 60335-1, IEC 60598-1, IEC 60884-1

Structural characteristics of the tracking tester leaks:
1. Combustion box paint treatment; Beautiful and elegant, corrosion resistant; Size: 1200mm It consists of control system, button switches, signal lights, etc. ;
4. The automatic drip device includes a drip cylinder, a drip nozzle, a counting device, a drip counting device, a drip control device, a drip volume measuring device, etc. ;
5. The drip nozzle device is adjustable; adjust the height by 30-40mm; make sure the dripping position is between the two electrodes;
6. The electrode devices includes platinum electrodes, electrode clamps, 4mm gauges, etc. The electrodes are adjustable and removable;
br>7. The electrode is 5mm wide, 2mm thick, 99.9% pure platinum, with a tip of 30-60 degrees;
8. The pressure of the electrode on the sample is adjustable, the adjustment range is (1.00N±0.001N); the electrode distance is 4.0mm ± 0.01mm, inclusive angle 60° ± 5°;
9. WithSample preparation step;
10. Equipped with glass plate: thickness ≥4mm;
11. Equipped with electrolyte: one bottle of ammonium chloride and sodium alkylnaphthalene sulfonate reagent;
Safety control system:
12. Test voltage: 0 ~ 1000 V, digital display continuously adjustable, maximum error of voltmeter display value: 1%;
13. Short circuit current: (1 ± 0.1) A, voltage drop value does not exceed 10%;
14. If the current of tis exceeds 0. If the duration exceeds 2 seconds, the system automatically stops the equipment;
15. Security door control system; When the door is open, the test system cannot work;
16. Electrode: 5mm × 2mm. One pair of rectangular platinum electrodes and one pair of brass electrodes, meeting the size requirements of IEC60112 Figure B.1;
17. Electrodes: The angle between the electrodes is 60° ± 5° and the spacing is (4 ± 0.1 mm);
18. Electrode: The pre-sample guarantee is: 1. 00N ± 0.05N;
19. Dripping system: (30 ± 5) seconds (starting dripping time 29.8S + dripping time starting drip 0.2S) automatic counting, digital display (can be preset), 50 droop time: (24.5 ± 2) min;
20. Drop system: Height of drop needle to sample surface: 35 mm ± 5 mm (a gauge is attached for reference);
21. Fall system: chu weightte: 20 drops: 0.380g ~ 0.489g; 50 drops: 0.997g ~ 1.147g;
22. PLC combined with touch screen control system makes the test more intelligent and more convenient to use;
23. Test data is automatically recorded, automatically stored and can be freely retrieved.