The bilirubin measurement radiometer is mainly used for visible light blue light phototherapy lamps or LED photovoltaic array testing. It can be used to test the intensity of bilirubin light, blue light LED, glasses photochemical blockage, aquarium lamps, acne lamps, and photosynthetic spectral blue bands, and has a wide range of applications.
Product Description:
The GaP photodiode is packaged in a UV glass window cover with a sealed gap.
The reduced response spectrum of Schott colored glass filter is shown in the response graph
Product operation:
Press and hold the switch button on the front of the solar radiometer. Align the sensor section at the top of the panel directly with the blue light source. The reading will be displayed on the LCD, and the testing distance will be recorded to ensure the accuracy of the repeated test results.
During operation, the battery voltage is between 9V-6.5V. When the voltage is below 6.5V, the LCD reading will start to blur, and the battery needs to be replaced. In "typical" mode, the battery can be used for about 2 years.
Product features:
This instrument contains an integrated sensor, which is easy to operate and durable. The test reading is accurate, and the data can be directly displayed on the LCD
Technical parameters:
Radiation range: 0-1999? W/cm?
633 nm response: 98%
Broadband: 422-499 nm
Resolution: 1? W/cm?
Update speed: 3.0 Readings/Sec
Display: 3.5 Digital LCD
Number size: 0.4 inch high
Operating humidity: 5% to 90% RH
Accuracy: ± 10% REF.NIST
Dimensions: 4.2L x 2.4W x 0.9D (in.)
Weight: 4.5 OZ. (included. bat.)
Power supply: 9-Volt DC Battery
Lens: UV Glass
Diffuser: Teflon
Detector: GaP with filter
Notes:
Eye protection is required when detecting blue light tubes
The front lamp tube needs to be illuminated for at least 5 minutes.
Keep the overall sensor of the instrument close to the light source to measure the intensity of individual lights or LEDs. To achieve effective intensity, maintain the working distance between the sensor and the light source.
When measuring the degree of aging of the lamp tube, the distance between the lamp tube and the testing meter needs to be kept constant.
When the light output of the lamp tube drops to about 70% of its original value, the lamp tube needs to be replaced.
If you are unsure of the original reading of the lamp, you can replace the old one with two new lamps of the same model and compare the readings.
Do not store the instrument in high temperature, damp, or dusty areas.
Wipe the instrument with a dry, soft cloth. Please keep away from grease and dust when storing.