Welcome to the Qinsun Instruments Co., LTD! Set to the home page | Collect this site
The service hotline

Search


Related Articles

Product Photo

Contact Us

Qinsun Instruments Co., LTD!
Address:NO.258 Banting Road., Jiuting Town, Songjiang District, Shanghai
Tel:021-67801892
Phone:13671843966
E-mail:info@standard-groups.com
Web:http://www.qinsun-lab.com

Your location: Home > Related Articles > Medical Mask Flame Retardant Performance Tester Instructions for Use أداة الاختبار –

Medical Mask Flame Retardant Performance Tester Instructions for Use أداة الاختبار –

Author:QINSUN Released in:2023-10 Click:107

The flame retardant performance tester of medical masks is mainly used to test the combustion performance of medical masks after contacting flames at a certain linear speed. This is a flame-retardant performance testing instrument for medical mask. This is a test equipment independently developed and produced by qisnun Precision Mechanical and Electrical Technology Co., Ltd.

Complies with standards:

GB19083-2010, YY0469-2011, GB2626-2006, BSEN136-1998

Usage Questions:

1. The test must be carried out under a fume hood.

2. Fireworks should be strictly prohibited in instrument and gas cylinder storage areas, and the fire should be eliminated. Samples that are still smoldering after testing should be extinguished promptly. Fire extinguishing equipment such as carbon dioxide should be installed.

3. The gasduc should be carefully checked before starting the machine and gas leakage is strictly prohibited.

4. Operators should wear protective clothing and gloves to prevent hands from being stabbed or burned.

5. The working room should be equipped with an exhaust fan to remove harmful gases and replace it with fresh air immediately after the test.

6. At the end of the test, be sure to turn off the gas source first, press and hold the \"ignition\" button, turn on the igniter and wait for the gas in the tube burns, then turn off the power.

7. If the igniter stops for any reason after \"starting\", please press the \"igniter return\" button. Once the igniter returns to the ignition position, press the “clear” button again. After resetting the working state, restart the operation.

8. If the air source is connected andthe “ignition” button is pressed, but the igniter fails to ignite, please check if the igniter port is blocked and remove all accessories.

As an enterprise integrating R&D, manufacturing, sales, training and service, Shanghai Fanbiao Textile Testing Technology Co., Ltd. is committed to continuously providing more testing instruments to the market and providing long-term textile supply. The testing instruments for leather, combustion, automobile interior materials, environmental climate aging, masks, etc. -standard customization. Targeted testing can be carried out according to customer requirements before purchase to ensure that the purchased instruments are suitable.

Prev: