As a laboratory instrument, the double-layered glass reactor should be used in strict accordance with the operating procedures. Now the common failure phenomena and causes of Double Layer Glass Reactor are sorted as follows:

Common Failure Phenomena of Double Layer Glass Reactor And Reasons:

1. Turn on the power switch, the light is not on: the external power supply is not connected or the contact is bad, please check the power supply and the socket by an electrician;

2. The fuse is shorted: turn off the power Turn the switch to the OFF position, then replace the fuse;

3. The power indicator is on , but does not rotate: The rotating shaft is rusty, stop using it and contact the supplier. The motor and electrical box are faulty, the seven-pin plug is not plugged in and the seven-pin plug is out of order.st reconnected;

4. Vacuum suddenly disappears: glass is cracked, switch is damaged, check glass parts, and replace switch;

5 .There is a vacuum, but it cannot be pumped: the sealing ring is worn and the vacuum interrupter connection is leaking, please replace the sealing ring and the interrupter;< p> 6. Sometimes there is no vacuum: there is dirt on the steel shaft and the connector is loose. Please remove the dirt and check the vacuum gauge and vacuum pump;

7. Vacuum hose aging: Please replace the vacuum hose.