Installation and commissioning manual DHM-200 ultrasonic height and weight scale

Author:QINSUN Released in:2023-04 Click:114

Luoyang Jiecan Electromechanical Co., Ltd. specializes in the production of
ultrasonic height and weight scales, computer scales, children\'s height and
weight scales, ultrasonic height and weight instruments, ultrasonic human health
scales, multifunctional height scales, ultrasonic voice broadcast scales,
portable ultrasonic height and weight scales, folding height - and weight
scales, height and weight measuring instruments, ultrasonic child seat height
scales, seat height and weight scales, baby height and weight scales, horizontal
height and weight scales.

Currently cattleModels used are:

DHM-200 (measure height, weight, BMI)

DHM-200D (measure height, weight, BMI, with imprint)

DHM- 200Y (measure height, weight, BMI, with seat)

DHM-300( Measure height, weight, BMI, foldable, type 301 with imprint)

DHM-600 (measure height, weight, BMI, with imprint)

DHM-600A (measure height, weight, BMI, with imprint, can measure blood
pressure and fat)

DHM-600B (measure height, weight, BMI, with imprint, blood pressure
measurement)

Mount machine:

DHM-16 (Economic version can measure height, weight, BMI, coin box,
printable)

DHM-15 (Deluxe version can measure height, weight, BMI, coin operated,
printable, measure)

DHM-15A (Deluxe version can measure height, weight, BMI, coin, pressure,
blood pressure, fat)

DHM-15B (Deluxe version can measure height, weight, BMI, coin box, printable,
measure blood pressure)

1. Date adjustment

Select menu code 1-01, press OK to enter the adjustment status .

The height window displays 0603, indicating that the current date is March
2006, and the weight window displays 24 again is the 24th. The blinking content
indicates that it is in the adjustment state.

Press + or - to adjust the displayed number to the date.

Press the selection cycle to change the year, month and day to be
modified

Press OK to save the adjustment result and exit. Press Cancel to exit without
saving the adjustment result.

2. Time adjustment

Select menu code 1-02, press OK to enter adjustment status .

The weight window displays 17:05 indicates the current time. The blinking
content indicates that it is in the adjustment state.

Press + or - to adjust the displayed digits to the actual time.

Press to select to cycle through minutes and hoursleaves to be modified.

Press OK to save the adjustment and exit. Press Cancel to exit without saving
the adjustment result.

3. Weight adjustment

Select menu code 1-03, Press OK to enter weight adjustment status.

The weight window shows 0.0 as the current weight. Place an object of known
weight (or your own weight) on the scale.

After the weight display is stable, press + or - to adjust the displayed
weight value to the actual weight.

If you need to measure again, you can press to select.

Press OK to save the customization and exitn. Press Cancel to exit without
saving the adjustment result.

4. Height adjustment

Place an object of known height on the weighing pan.

Select the menu code 1-04, press OK to enter the height setting and start
measuring.

The height window displays 170.3 as the current altitude.

After the altitude display is stable, press + or - to adjust until the
displayed altitude value matches the actual altitude.

When measuring again, you can press Select.

Press OK to save and print the adjustment result to close. Press Cancel to
exit without saving the adjustment result.

5.Inactive speech interval setting:

Select menu code 1- 05, press OK to enter interval adjustment status.

The weight window shows 6 as the current parameter, and the range is 0 -50,
which means that the interval of reporting \"Welcome\" is 30 seconds.

Press + or - to adjust the displayed number to the desired value.

Press OK to save the adjustment results and exit. Press Cancel to exit
without the adjustment results.

Note: a. Actual time interval = set value*5 (unit : second).

b. When the set trueis 0, \"Welcome\" is not reported.

6. Voice status setting:

Select menu code 1-06, press OK Enter the adjustment status.

Weight window shows 01 range 00-07.

00: no vote; 01: report height and weight; 02: report body shape; 03: report
height, weight and body shape ; 04: static and rich voice; 05: report height and
weight, static rich voice; 06: report body shape, static rich voice; 07: Report
height, weight, body shape, static rich voice.

Press + or - to adjust the displayed number to the desired value.

Press OK to save the customization and exit. Press Cancel to exitclose e
without saving the adjustment result.

7. BMI normal range maximum:

Choose menu code 1-07 >, press OK to access the customization status.

The weight window shows 25.0 is the maximum body mass index, ranging from
19.0-28.0.

Press + or - to adjust the displayed number to the desired value.

Press OK to save the adjustment and exit. Press Cancel to exit without saving
the adjustment results.

8. Minimum BMI normal range:

Choose menu code 1-08, press OK to enter the adjustment statusto go.

The weight box indicates that 19.0 is the maximum body mass index and the
range is 16.0- 22.0.

Press + or - to adjust the displayed number to the desired value.

Press OK to save the adjustment and exit. Press Cancel to exit without saving
the adjustment results.

Note: a. The range of body mass index in different regions may differ and the
sub-items should be adjusted according to actual requirements.

b. The maximum and minimum values ​​of the body mass index determine the
display position of the body shape light after measurement and ideal weight and
standard weight range for print measurements.

9 . Print mode setting: (for models 200D/301/16)

Select menu code 1- 09, press OK to enter adjustment status.

The weight box shows that 01 is the current print parameter and the range is
0-8. See table below for parameter meanings.

Press + or - to adjust the displayed number to the desired value.

Press OK to save the adjustment and exit. Press Cancel to exit without saving
the adjustment results.