Taber abrasion instrument operation

Author:QINSUN Released in:2021-03 Click:582

Taber abrasion tester is suitable for cloth, paper, paint, plywood, leather, floor tiles, glass, natural plastic, etc. The test method is that the rotating sample bears a pair of abrasion wheels, and the specified load is applied. When the sample rotates, the abrasion wheel is driven to wear the sample. The abrasion loss weight is the difference between the weight of the sample before and after the test.

Taber abrasion instrument operation

Technical Parameters:

  Equipment model: T270

  1, test piece: inner diameter (D) 3mm

  2, wear wheel: ф2"(Max.45mm)(W)1/2"

  3, wear wheel center distance: 63.5mm

  4. The distance between the wear wheel and the center of the test plate: 37~38mm

  5, rotation speed: 60r/min

   6. Load: 250, 500, 1000g

  7, counter: LED 0~99999999

   8. The distance between the test piece and the dust suction port: 3mm

  9, volume: 45×35×42cm

  10, weight: 28kg

  11, power supply: 1∮AC 220V, 10A

  12. Random configuration: a pair of test wheels and a vacuum cleaner

Standards compliant:

  DIN53754,53799,53109,52347,TAPPI T476, ASTM D3884,ISO 54701,ASTM D 1175

   Test preparation:

   1. Fix the sample on the test turntable, select the required test weight and install it, and make sure that the test wheel does not touch the sample fixture during the test.

   2. Turn the governor switch to the lowest.

   3. After setting the test times according to the test requirements, the whole machine is in the test state.

  Operation Specifications:

   1. Connect the appropriate power 220V or 110V according to the label

   2. Turn the governor switch to the lowest level to avoid excessive speed, which may cause abnormalities in the machine.

   3. Turn on the power, test the machine first, and turn off the power if there is no abnormality.

   4. Press the reset button to reset the counter to zero. Set the required test times according to the requirements, and the counter can be set up to 99999999 times. Press the RST key on the counter to clear the counter, and press the> key for five seconds to enter the test number setting. The eight digits on the counter from the lower right to the upper left represent the ones digit to the ten million digits. If you set 280,360 times, press the> key. Adjust the beating value to the third digit from the upper left and press the ∧ key to set the value to 2, then press the> key to adjust the beating value to the fourth digit from the upper left, press the ∧ key to set the value to 8, and then press the> key to jump Adjust the value to the third digit from the bottom right and press the ∧ key to set the value to 3, then press the> key to adjust the beating value to the second digit at the bottom right and press the ∧ key to set the value to 6, and all other positions are set to 0 is enough,

  5. Set the required number of tests according to the test requirements. Upper left one upper left two upper left three upper left four lower right four lower right three lower right two lower right

   6. Press the start button, the whole machine will enter the test state.

   7. Adjust the governor to adjust the speed to the required test speed.

   8. Turn on the power switch of the vacuum cleaner to make it work.

   9. After the number of times set by the counter is tested, the machine will automatically stop and be in the standby state.

   10. Turn off the power, remove the sample, and clean the machine.

   Maintenance items:

   1. The instrument must be placed on a stable ground before use,

   2. It is not allowed to move the machine during use;

  3. Choose the corresponding power supply voltage, do not be too high, to avoid burning the device;

  4. When the instrument is abnormal, please contact our company for timely handling;

   5. There must be a good ventilation environment when the machine is working.

  6. ​​Due to the large operating load of each component of the button mechanism, please add lubricating oil to the mechanical part in time;

  7. After each test, clean the machine and keep the machine clean;

   8. The control box part must be wiped with a dry cloth, not a wet cloth.