One Article Understands Taber Abrasion Tester

Author:QINSUN Released in:2017-06 Click:1402

Taber abrasion testing machine mainly through the rotation of the grinding method, the use of different types of friction wheel on the sample rotation to detect leather and plastic materials such as wear resistance.
Test preparation:
1. The test specimen is fixed on the test turntable, select the required test weight loaded;
2. Ensure that the test wheel does not come into contact with the sample fixture during the test;
3. Adjust the governor switch to a minimum;
4. According to the test requirements set the number of tests, the machine is in the test state;

Code of Practice:
1. Connect the appropriate power supply 220V or 110V;
2. adjust the governor switch to a minimum, in order to avoid the speed too fast, resulting in abnormal machine;
3. Turn on the power, first test machine, if no abnormal power is turned off;
4. Press the reset button to reset the counter to zero. According to the requirements set to test the number of tests, the maximum number of counters can be set 99999999 times;
5. Press the RST key on the counter to clear the counter;
6. Press the start button, the machine will enter the test state;
7. Adjust the governor to adjust the speed to the desired test speed;
8. Turn on the vacuum cleaner power switch to start working;
9. Test to the counter set the number of times, the machine automatically shut down in standby mode;
10. Turn off the power, remove the test, clean up the machine;

Maintenance matters:
1. The instrument must be placed on a smooth ground before use;
2. Do not allow the use of the process of naked machine;
3. Select the appropriate power supply voltage, not too high, to avoid burning device;
4. When the instrument is abnormal, please contact with the Division for timely processing;
5. Machine work must have a good ventilation environment;
6. key mechanical components, due to the operation of a larger load, the mechanical part of the timely increase of lubricants;
7. After each test is completed, clean up the machine, keep the machine clean;

Common malfunctions:
1. If the power is turned on, it indicates that there is no power on or the power indicator is damaged. If the power indicator is damaged, replace the indicator light. Such as the power output terminal no power output Please check the power cord or power outlet;
2. If the power is on, the power indicator has an indication. Press the start button. The machine does not work. Please check whether the relay has dropped or disconnected, such as dropped or disconnected. If the relay is disconnected, , Indicating that the relay is damaged, it should be replaced;
(1) the counter is damaged; (2) after the counter is set, it is not cleared; after the counter is set to the number of times, the motor will not stop working,
4. If the power indicator is on, the counter is not displayed, indicating that the counter is damaged, you need to replace the counter;